WAN J H, GUO C J, ZHU H. The design,calculation and adjustment of multi-differential vacuum system[J]. Journal of South China Normal University (Natural Science Edition), 2015, 47(4): 35-40. DOI: 10.6054/j.jscnun.2014.12.054
Citation: WAN J H, GUO C J, ZHU H. The design,calculation and adjustment of multi-differential vacuum system[J]. Journal of South China Normal University (Natural Science Edition), 2015, 47(4): 35-40. DOI: 10.6054/j.jscnun.2014.12.054

The design,calculation and adjustment of multi-differential vacuum system

  • Linear ion trap-time of flight mass spectrometer (LIT-TOFMS) is one of multi-stage tandem mass spectrometry with high detection efficiency, high resolving power and high sensitivity. The differential vacuum system is the foundation of LIT-TOFMS. Basing on the principle of differential vacuum system, this paper introduces the choice of materials, vacuum measurement elements and pumping system. By the calculation, the LIT-TOFMS vacuum system- four-stage differential vacuum system is successfully designed. Both the calculations and the experiment results can prove that this system could meet the requirements of LIT-TOFMS and realize the pressure transition from atmospheric pressure to 10-4Pa, which is a reliable guarantee for the research and development of LIT-TOFMS.